Please use this identifier to cite or link to this item: https://hdl.handle.net/2440/87963
Citations
Scopus Web of Science® Altmetric
?
?
Full metadata record
DC FieldValueLanguage
dc.contributor.authorDing, L.-
dc.contributor.authorYang, G.-
dc.contributor.authorXia, F.-
dc.contributor.authorLenehan, C.-
dc.contributor.authorQian, G.-
dc.contributor.authorMcFadden, A.-
dc.contributor.authorBrugger, J.-
dc.contributor.authorZhang, X.-
dc.contributor.authorChen, G.-
dc.contributor.authorPring, A.-
dc.date.issued2011-
dc.identifier.citationMineralogical Magazine, 2011; 75(2):279-287-
dc.identifier.issn0026-461X-
dc.identifier.issn1471-8022-
dc.identifier.urihttp://hdl.handle.net/2440/87963-
dc.description.abstractThe accurate measurement of trace element concentrations in natural sulphides by laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) has been limited by the lack of matrix-matched calibration standards. The synthesis of a standard, IMER-1, by incorporating four minor and 34 trace elements into a chalcogenide glass matrix Ge28Sb12S60 is reported here. Chemical analysis by electron probe microanalysis (EPMA), LA-ICP-MS, solution ICP-MS, and inductively coupled plasma-optical emission spectroscopy (ICP-OES) confirmed the excellent homogeneity of major elements (1-σ relative standard deviation (RSD) <1% for S, Sb and Ge) and acceptable homogeneity of most trace elements (1-σ RSD <10%). The standard was validated by analysing trace-elements concentrations in three geological pyrite specimens using IMER-1 as the calibration standard and comparing the results to previously reported values also determined by LA-ICP-MS but using a different calibration standard, STDGL2b-2. The results suggest that IMER-1 may be an appropriate calibration standard for LA-ICP-MS analysis of trace elements in natural sulphides.-
dc.description.statementofresponsibilityLihua Ding, Guang Yang, Fang Xia, Claire E. Lenehan, Gujie Qian, Aoife McFadden, Joël Brugger, Xianghua Zhang, Guorong Chen, Allan Pring-
dc.language.isoen-
dc.publisherMineralogical Society-
dc.rights© 2011 The Mineralogical Society-
dc.source.urihttp://dx.doi.org/10.1180/minmag.2011.075.2.279-
dc.subjectGe-Sb-S chalcogenide glasses; LA-ICP-MS; sulphide calibration standard; trace element analysis-
dc.titleA LA-ICP-MS sulphide calibration standard based on a chalcogenide glass-
dc.typeJournal article-
dc.identifier.doi10.1180/minmag.2011.075.2.279-
pubs.publication-statusPublished-
Appears in Collections:Aurora harvest 7
Earth and Environmental Sciences publications

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.