A validation study of a commercial wearable device to automatically detect and estimate sleep

Date

2021

Authors

Miller, D.J.
Roach, G.D.
Lastella, M.
Scanlan, A.T.
Bellenger, C.R.
Halson, S.L.
Sargent, C.

Editors

Advisors

Journal Title

Journal ISSN

Volume Title

Type:

Journal article

Citation

Biosensors, 2021; 11(6, article no. 185):1-14

Statement of Responsibility

Conference Name

Abstract

The aims of this study were to: (1) compare actigraphy (ACTICAL) and a commercially available sleep wearable (i.e., WHOOP) under two functionalities (i.e., sleep auto-detection (WHOOP-AUTO) and manual adjustment of sleep (WHOOP-MANUAL)) for two-stage categorisation of sleep (sleep or wake) against polysomnography, and; (2) compare WHOOP-AUTO and WHOOP-MANUAL for four-stage categorisation of sleep (wake, light sleep, slow wave sleep (SWS), or rapid eye movement sleep (REM)) against polysomnography. Six healthy adults (male: n = 3; female: n = 3; age: 23.0 ± 2.2 yr) participated in the nine-night protocol. Fifty-four sleeps assessed by ACTICAL, WHOOP-AUTO and WHOOP-MANUAL were compared to polysomnography using difference testing, Bland–Altman comparisons, and 30-s epoch-by-epoch comparisons. Compared to polysomnography, ACTICAL overestimated total sleep time (37.6 min) and underestimated wake (−37.6 min); WHOOP-AUTO underestimated SWS (−15.5 min); and WHOOP-MANUAL underestimated wake (−16.7 min). For ACTICAL, sensitivity for sleep, specificity for wake and overall agreement were 98%, 60% and 89%, respectively. For WHOOP-AUTO, sensitivity for sleep, wake, and agreement for two-stage and four-stage categorisation of sleep were 90%, 60%, 86% and 63%, respectively. For WHOOP-MANUAL, sensitivity for sleep, wake, and agreement for two-stage and four-stage categorisation of sleep were 97%, 45%, 90% and 62%, respectively. WHOOP-AUTO and WHOOP-MANUAL have a similar sensitivity and specificity to actigraphy for two-stage categorisation of sleep and can be used as a practical alternative to polysomnography for two-stage categorisation of sleep and four-stage categorisation of sleep.

School/Discipline

Dissertation Note

Provenance

Description

Access Status

Rights

Copyright 2021 The author(s). This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited (https://creativecommons.org/licenses/by/4.0/)

License

Call number

Persistent link to this record