Simplified three-cornered-hat technique for frequency stability measurements

Date

2014

Authors

Gunn, L.
Catlow, P.
Al-Ashwal, W.
Hartnett, J.
Allison, A.
Abbott, D.

Editors

Advisors

Journal Title

Journal ISSN

Volume Title

Type:

Journal article

Citation

IEEE Transactions on Instrumentation and Measurement, 2014; 63(4):889-895

Statement of Responsibility

Lachlan J. Gunn, Peter G. Catlow, Waddah A. Al-Ashwal, John G. Hartnett, Andrew Allison, and Derek Abbott.

Conference Name

Abstract

We propose a novel technique allowing the use of the three-cornered-hat method with two devices under test (DUTs) and a time-tagging system that employs a common reference oscillator. The precision of a time-tagging system is reduced by fluctuations in the timebase, which are canceled when the relative phase between the two DUTs is measured. However, the raw time-tags in this system provide a phase comparison between the DUTs and the system timebase, allowing the use of the three-cornered hat with some dual-channel measurement instruments.