Early clinical evaluation of the Hugo robotic-assisted surgery (RAS) for performing radical prostatectomy: an IDEAL stage 2 study

Files

hdl_146785.pdf (314.17 KB)
  (Published version)

Date

2025

Authors

Shepherd, A.
Jaffer, A.
Bruce, A.
Chia, D.
Dasgupta, P.
Challacombe, B.

Editors

Advisors

Journal Title

Journal ISSN

Volume Title

Type:

Journal article

Citation

BMJ Surgery, Interventions, & Health Technologies, 2025; 7(1):e000360-1-e000360-5

Statement of Responsibility

Andrew Shepherd, Ata Jaffer, Angus Bruce, Daniel Chia, Prokar Dasgupta, Ben Challacombe

Conference Name

Abstract

Objectives: To assess the feasibility and safety of the new Hugo robotic-assisted surgery (RAS) system for robotic-assisted radical prostatectomy (RARP), describing iterative changes in our operative technique—IDEAL stage 2. Design: Prospective, single-centre series. Setting: Tertiary urological unit in London, UK. Participants: Male patients diagnosed with clinically localised prostate cancer and suitable for RARP from February 2023 to May 2024. Main outcome measures: The primary outcome was to assess the safety of using the device without converting to the existing robotic platform (da Vinci), laparoscopy or open. Secondary outcomes assessed surgical (operative time, blood loss, time to catheter removal, complications), oncologic (surgical pathology and margin status) and early functional (continence) outcomes. Results: 50 patients were included in the study. No cases required conversion to an existing robotic platform, laparoscopy or open, and there were no intraoperative surgical complications. Mean age was 60 years and mean prostate-specific antigen was 12.2 ng/mL. The mean operative time was 148 min and estimated blood loss was 168 mL. Mean length of stay was 1.5 days and mean length of catheter duration was 13 days. On final pathology, 18 patients (36%) had T3 disease and four had positive surgical margins (8%). The mean International Consultation on Incontinence Questionnaire–Urinary Incontinence score for urinary continence at 3 months was 7. There were six Clavien-Dindo grade 2 complications and two Clavien-Dindo 3a complications. There were four instances of recoverable, temporary device failure. Iterative improvements were made to docking setup, use of robotic instruments and reduction in robotic arm collisions. Conclusions: We demonstrated feasibility and the safe introduction of the Hugo RAS for RARP into an experienced robotic urological programme. Perioperative, early oncological and functional outcomes were similar to other early series. Further studies will aim to describe the learning curve with this robot and optimisation of surgical quality.

School/Discipline

Dissertation Note

Provenance

Description

Access Status

Rights

© Author(s) (or their employer(s)) 2025. Re-use permitted under CC BY. Published by BMJ Group. Open access. This is an open access article distributed in accordance with the Creative Commons Attribution 4.0 Unported (CC BY 4.0) license, which permits others to copy, redistribute, remix, transform and build upon this work for any purpose, provided the original work is properly cited, a link to the licence is given, and indication of whether changes were made. See: https://creativecommons.org/licenses/by/4.0/.

License

Grant ID

Call number

Persistent link to this record