Characterization of the electrochemical behavior of gastrointestinal fluids Using a multielectrode sensor probe

Date

2011

Authors

Twomey, K.
Alvarez de Eulate, E.
Marchesi, J.R.
Kolida, S.
Gibson, G.
Arrigan, D.W.M.
Ogurtsov, V.I.

Editors

Advisors

Journal Title

Journal ISSN

Volume Title

Type:

Journal article

Citation

IEEE Transactions on Biomedical Engineering, 2011; 58(9):2521-2527

Statement of Responsibility

Conference Name

Abstract

A characterization of gastrointestinal fluids has been performed by means of an electrochemical sensor that has potential for clinical in vivo and in vitro monitoring applications. The sensor comprised a three-electrode cell with a counter, reference, and four working electrodes, Au, Pt, Ir, and Rh. Cyclic voltammetry was used to obtain chemical information from faecal water (in vitro) and gut model (in vivo) fluids. Stable voltammetric responses were obtained for both fluids at these noble metal working electrodes. The responses differed in shape that demonstrated the discrimination capability and the potential for practical use as a tool for gastrointestinal fluid investigation. The analysis of the stability profiles in faecal water over a 14-h duration has indicated a possible adsorption mechanism with the formation of a biolayer on the sensor surface. The stability in gut model fluids over a 42-h duration has demonstrated a more stable profile, but the mechanisms involved are more complicated to determine.

School/Discipline

Dissertation Note

Provenance

Description

Link to a related website: https://espace.curtin.edu.au/bitstream/20.500.11937/26790/2/168824_168824.pdf, Open Access via Unpaywall

Access Status

Rights

Copyright 2011 IEEE

License

Grant ID

Call number

Persistent link to this record