Takam, R.Bezak, E.Liu, G.Marcu, L.2012-10-152012-10-152012Radiation Protection Dosimetry, 2012; 150(1):22-330144-84201742-3406http://hdl.handle.net/2440/73576⁶LiF:Mg,Cu,P and ⁷LiF:Mg,Cu,P glass-rod thermoluminescent dosemeters (TLDs) were used for measurements of out-of-field photon and neutron doses produced by Varian iX linear accelerator. Both TLDs were calibrated using 18-MV X-ray beam to investigate their dose–response sensitivity and linearity. CR-39 etch-track detectors (Luxel+, Landauer) were employed to provide neutron dose data to calibrate 6LiF:Mg,Cu,P TLDs at various distances from the isocentre. With cadmium filters employed, slow neutrons (<0.5 eV) were distinguished from fast neutrons. The average in-air photon dose equivalents per monitor unit (MU) ranged from 1.5±0.4 to 215.5±94.6 μSv at 100 and 15 cm from the isocentre, respectively. Based on the cross-calibration factors obtained with CR-39 etch-track detectors, the average in-air fast neutron dose equivalents per MU range from 10.6±3.8 to 59.1±49.9 μSv at 100 and 15 cm from the isocentre, respectively. Contribution of thermal neutrons to total neutron dose equivalent was small: 3.1±7.2 μSv per MU at 15 cm from the isocentre.en© The Author 2011FluoridesLithium CompoundsGlassEquipment DesignEquipment Failure AnalysisRadiation DosageThermoluminescent DosimetryDose-Response Relationship, RadiationParticle AcceleratorsThe use of enriched ⁶Li and ⁷Li Lif:Mg,Cu,P glass-rod thermoluminescent dosemeters for linear accelerator out-of-field radiation dose measurementsThe use of enriched (6)Li and (7)Li Lif:Mg,Cu,P glass-rod thermoluminescent dosemeters for linear accelerator out-of-field radiation dose measurementsJournal article002012002410.1093/rpd/ncr3590003048359000022-s2.0-8486231320224198Bezak, E. [0000-0002-1315-1735] [0000-0003-0084-4089]