Chau, A.Fung, K.Yap, M.2025-12-182025-12-182005Radiography, 2005; 11(1):35-391078-8174https://hdl.handle.net/11541.2/143959Purpose: This study reports a calibration carried out on phantoms simulating the orbit and eyeball to evaluate the accuracy of volumetric determination using MRI DICOM (Digital Imaging and Communication in Medicine) images. Methods: Ten tailor-made spherical silicon balls ranging in size from 5 to 14 cm3 and 10 silicon moulds of orbits from 10 human dried skulls were used to simulate the eyes and orbits, respectively. The CISS (Constructive interference in steady state, TR/TE = 16/8 ms) T2-weighted sequence was taken using a Siemens MRI scanner. The volume of each phantom was computed and compared with the known physical volumes. Results: The computed and physical volumes were highly correlated for both eyeball (r = 0.997) and the orbit (r = 0.992) phantoms. Coefficients of variation of the computed and physical volumes were low. Consequently, it is possible to apply a calibration value to the computed volume to derive the physical volume with a high level of confidence. Conclusion: We conclude that with appropriate calibration, it is viable to use MRI DICOM images to derive the volume of the eyeball and the orbit.enCopyright 2004 The College of Radiographerseyesmagnetic resonance imagingimage analysisEvaluation of the accuracy of volume determination on the orbit and eyeball using MRIJournal article10.1016/j.radi.2004.10.0042-s2.0-12844273862