Investigation of a MOSFET dosimetry system for midpoint dose verification in prostate 3D CRT/IMRT

Date

2008

Authors

Wiese, T.
Bezak, E.
Nelligan, R.

Editors

Advisors

Journal Title

Journal ISSN

Volume Title

Type:

Journal article

Citation

Australasian Physical and Engineering Sciences in Medicine, 2008; 31(3):180-190

Statement of Responsibility

Wiese, T; Bezak, E and Nelligan, R

Conference Name

Abstract

The suitability of MOSFETs (Metal Oxide Semiconductor Field Effect Transistors) for use in in-vivo dosimetry for IMRT prostate treatment and patient setup errors has been investigated in this work. MOSFETs were placed on entrance and exit surfaces of a number of different phantoms (with varying complexities from homogeneous to anthropomorphic). Dose measurements were then used to calculate a midpoint dose, which was compared with an IC placed at the isocentre. The agreements found between the calculated (MOSFETs) and the measured midpoint dose (IC) was: 0.7 % for a prostate treatment verification and 3.5 % for an IMRT treatment. MOSFETs placed on entry and exit surfaces can detect patient setup offsets of 2 cm, but do not have the sensitivity to confidently detect offsets of 1 cm or smaller.

School/Discipline

Dissertation Note

Provenance

Description

Access Status

Rights

Copyright 2008 ACPSEM

License

Grant ID

Call number

Persistent link to this record