High temperature liquid chromatography with monolithic capillary columns and pure water eluent
Date
2009
Authors
Causon, T.J.
Shellie, R.A.
Hilder, E.
Editors
Advisors
Journal Title
Journal ISSN
Volume Title
Type:
Journal article
Citation
The Analyst, 2009; 134(3):440-442
Statement of Responsibility
Conference Name
DOI
Abstract
Use of an organic polymeric monolithic stationary phase for high temperature liquid chromatography (HTLC) with a pure water eluent is demonstrated for the first time. Separations of n-alcohols at temperatures exceeding 200°C were achieved with no statistically significant reduction in chromatographic performance evident after continuous operation for 1000 column volumes above 200°C. © The Royal Society of Chemistry.
School/Discipline
Dissertation Note
Provenance
Description
Access Status
Rights
Copyright 2009 Royal Society of Chemistry