High-speed cannula detachment into the eye during hydrodissection

dc.contributor.authorMcPherson, Z.E.
dc.contributor.authorLau, O.C.
dc.contributor.authorChen, T.S.
dc.contributor.authorKam, A.W.
dc.contributor.authorAmjadi, S.
dc.contributor.authorZhang, M.G.
dc.contributor.authorPlayfair, T.J.
dc.contributor.authorAgar, A.
dc.contributor.authorFrancis, I.C.
dc.date.issued2014
dc.description.abstractDetachment of a hydrodissection cannula during a phacoemulsification procedure appeared to produce no adverse sequelae during surgery. Day 1 postoperatively, two nonpenetrating hemorrhagic retinal lesions were identified; there was no evidence of posterior capsular perforation. Day 6 postoperatively, the pupil was temporally peaked by a fine vitreous strand running to the main-port incision in the superotemporal cornea. This was divided with Nd:YAG laser, and argon laser was applied to encircle the two retinal lesions. Postoperative uncorrected visual acuity remained 6/4 at day 1, day 6, and week 4 (3 weeks after laser application) follow-up visits. Surgeons must accept responsibility for confirming the integrity of the cannula and syringe connection before beginning hydrodissection, which can be highly destructive to intracameral structures.
dc.description.statementofresponsibilityZachary E. McPherson, Oliver C.F. Lau, Tony S. Chen, Andrew W. Kam, Shahriar Amjadi, Michael G. Zhang, T. Justin Playfair,Ashish Agar, Ian C. Francis
dc.identifier.citationOphthalmic Surgery, Lasers and Imaging, 2014; 45(4):347-349
dc.identifier.doi10.3928/23258160-20140624-01
dc.identifier.issn1938-2375
dc.identifier.issn2325-8179
dc.identifier.urihttp://hdl.handle.net/2440/93955
dc.language.isoen
dc.publisherSlack
dc.rightsCopyright status unknown
dc.source.urihttps://doi.org/10.3928/23258160-20140624-01
dc.subjectRetinal Hemorrhage
dc.titleHigh-speed cannula detachment into the eye during hydrodissection
dc.typeJournal article
pubs.publication-statusPublished

Files