Multispectral camera system design for replacement of hyperspectral cameras for detection of aflatoxin B1

Date

2022

Authors

Ariza Ramirez, W.
Mishra, G.
Panda, B.K.
Jung, H.W.
Lee, S.H.
Lee, I.
Singh, C.B.

Editors

Advisors

Journal Title

Journal ISSN

Volume Title

Type:

Journal article

Citation

Computers and Electronics in Agriculture, 2022; 198(107078)

Statement of Responsibility

Conference Name

Abstract

The design of a multispectral camera system to replace hyperspectral cameras used in chemometrics laboratory for food research is a challenging task in the field of machine vision. In this paper, the process of designing and prototyping a low cost multispectral camera to measure aflatoxin B1 in almonds is described. A multispectral system with a Goldeye G-008 TEC1 SWIR is developed to replace a hyperspectral system with a Specim FX17 line scanner. Using R-squared and RMSE values of multiple linear regression, the performance of a hyperspectral camera system with a multispectral camera system were compared for measuring aflatoxin B1. The results show that a high speed multispectral system can replace an equivalent hyperspectral imaging system with much lower cost with an acceptable processing performance of 20 datacubes per seconds.

School/Discipline

Dissertation Note

Provenance

Description

Access Status

Rights

Copyright 2022 Elsevier Access Condition Notes: Accepted manuscript available after 1 July 2024

License

Grant ID

Call number

Persistent link to this record