Wear of a highly cross-linked polyethylene liner: a preliminary RSA study

dc.contributor.authorCampbell, D.
dc.contributor.authorMercer, G.
dc.contributor.authorNilsson, G.
dc.contributor.authorWells, V.
dc.contributor.authorField, J.
dc.contributor.authorCallary, S.
dc.date.issued2009
dc.description.abstractWear induced peri-prosthetic osteolysis and aseptic loosening remain the major contributing factors to failure of total hip arthroplasty. To reduce wear particles from acetabular liners, the process of polyethylene cross-linking has been modified. In this study, we examine the wear of Marathon™ acetabular liners using Radiostereometric Analysis. Thirty patients were enrolled in a prospective cohort study. Femoral head penetration was calculated for 25 patients at 6 months, 1 and 2 years postoperatively. The mean proximal head penetration was 0.067 mm at 6 months, 0.113 mm at 1 year and 0.120 mm at 2 years. The mean three-dimensional femoral head penetration was 0.240 mm at 6 months, 0.230 mm at 1 year and 0.232 mm at 2 years. For both parameters, there was no measurable wear between 1 and 2 years, and the mean annual linear wear was less than 0.06 mm/year, which is less than the suggested osteolysis threshold for polyethylene. All patients demonstrated a dramatic improvement in clinical scores. These results were comparable with other studies that used alternative techniques with conventional radiography to evaluate wear for Marathon™. Long-term studies using the RSA method are recommended to further quantify the clinical performance of this polyethylene liner.
dc.description.statementofresponsibilityDavid Campbell, Graham Mercer, Kjell-Gunnar Nilsson, Vanessa Wells, John R. Field and Stuart A. Callary
dc.identifier.citationEuropean Journal of Orthopaedic Surgery and Traumatology, 2009; 20(1):23-27
dc.identifier.doi10.1007/s00590-009-0486-y
dc.identifier.issn1633-8065
dc.identifier.issn1432-1068
dc.identifier.orcidCampbell, D. [0000-0002-1572-9529]
dc.identifier.orcidCallary, S. [0000-0002-2892-5238]
dc.identifier.urihttp://hdl.handle.net/2440/57168
dc.language.isoen
dc.publisherSpringer France
dc.source.urihttps://doi.org/10.1007/s00590-009-0486-y
dc.subjectHip
dc.subjectWear
dc.subjectRSA
dc.subjectPolyethylene
dc.subjectRadiostereometric analysis
dc.subjectCross-linked
dc.titleWear of a highly cross-linked polyethylene liner: a preliminary RSA study
dc.typeJournal article
pubs.publication-statusPublished