Combined effect of cup abduction and anteversion angles on long-term wear evolution of PCD-on-PCD hip bearing couple

Date

2017

Authors

Shankar, S.
Gowthaman, K.
Uddin, M.

Editors

Advisors

Journal Title

Journal ISSN

Volume Title

Type:

Journal article

Citation

International Journal of Biomedical Engineering and Technology, 2017; 24(2):169-183

Statement of Responsibility

Conference Name

Abstract

This paper presents a three-dimensional (3D) finite element (FE) modelling for the estimation of wear evolution on the acetabulum cup against the femoral head for the combinations of different abduction angle (35°, 45° and 55°) and anteversion angle (0°, 10°, 20°, 30° and 40°) for PCD-on-PCD hip bearing components. By implementing the actual physiological hip gait loading conditions and the rotational movement of a normal walking cycle in the FE model, the contact pressure and the sliding distance were calculated to predict the wear. The cup surface was updated in a suitable interval of gait cycles for the estimation of wear. Wear was simulated for 20 years implant's life in a human body. Simulation results showed that higher abduction angle and lower anteversion angle result in the least linear and volumetric wear as compared to other conditions considered.

School/Discipline

Dissertation Note

Provenance

Description

Access Status

Rights

Copyright 2017 Inderscience Enterprises Ltd

License

Grant ID

Call number

Persistent link to this record