Surgical removal of duct occluder device under mild hypothermia without cardiopulmonary bypass

Date

2010

Authors

Mohite, Prashant N.
Kuthe, Sachin A.
Thingnam, Shyam K.
Rohit, Manoj K.
Mahajan, Sachin
Mahajan, Rajiv

Editors

Advisors

Journal Title

Journal ISSN

Volume Title

Type:

Journal article

Citation

Innovations: Technology & Techniques in Cardiothoracic & Vascular Surgery, 2010; 5(4):311-312

Statement of Responsibility

Prashant N. Mohite, Sachin A. Kuthe, Shyam K. Thingnam, Manoj K. Rohit, Sachin Mahajan and Rajiv Mahajan

Conference Name

Abstract

Because the use of percutaneous intervention is increasing for the closure of the patent ductus arteriosus, the procedure related complications are also on rise, with migration of the device being most common. The routine practice is to remove the migrated duct occluder device under cardiopulmonary bypass. Amplatzer duct occluder used in a 4-month-old infant dislodged into the descending thoracic aorta. It was removed by the posterolateral thoracotomy under mild hypothermia through juxtaductal aortotomy between the aortic cross-clamps. The use of cardiopulmonary bypass is thus avoided.