A longitudinal evaluation of performance of automated BCR-ABL1 quantitation using cartridge-based detection system

Files

hdl_101525.pdf (201.88 KB)
  (Published version)

Date

2015

Authors

Enjeti, A.
Granter, N.
Ashraf, A.
Fletcher, L.
Branford, S.
Rowlings, P.
Dooley, S.

Editors

Advisors

Journal Title

Journal ISSN

Volume Title

Type:

Journal article

Citation

Pathology, 2015; 47(6):570-574

Statement of Responsibility

Anoop Enjeti, Neil Granter, Asma Ashraf, Linda Fletcher, Susan Branford, Philip Rowlings, Susan Dooley

Conference Name

Abstract

An automated cartridge-based detection system (GeneXpert; Cepheid) is being widely adopted in low throughput laboratories for monitoring BCR-ABL1 transcript in chronic myelogenous leukaemia. This Australian study evaluated the longitudinal performance specific characteristics of the automated system.The automated cartridge-based system was compared prospectively with the manual qRT-PCR-based reference method at SA Pathology, Adelaide, over a period of 2.5 years. A conversion factor determination was followed by four re-validations. Peripheral blood samples (n = 129) with international scale (IS) values within detectable range were selected for assessment. The mean bias, proportion of results within specified fold difference (2-, 3- and 5-fold), the concordance rate of major molecular remission (MMR) and concordance across a range of IS values on paired samples were evaluated.The initial conversion factor for the automated system was determined as 0.43. Except for the second re-validation, where a negative bias of 1.9-fold was detected, all other biases fell within desirable limits. A cartridge-specific conversion factor and efficiency value was introduced and the conversion factor was confirmed to be stable in subsequent re-validation cycles. Concordance with the reference method/laboratory at >0.1-≤10 IS was 78.2% and at ≤0.001 was 80%, compared to 86.8% in the >0.01-≤0.1 IS range. The overall and MMR concordance were 85.7% and 94% respectively, for samples that fell within ± 5-fold of the reference laboratory value over the entire period of study.Conversion factor and performance specific characteristics for the automated system were longitudinally stable in the clinically relevant range, following introduction by the manufacturer of lot specific efficiency values.

School/Discipline

Dissertation Note

Provenance

Description

Access Status

Rights

Copyright © 2015 Royal College of Pathologists of Australasia. All rights reserved. This is an openaccess article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License, where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially. DOI: 10.1097/PAT.0000000000000293

License

Grant ID

Call number

Persistent link to this record