Design of a wireless power supply receiver for biomedical applications
Files
(Published version)
Date
2006
Authors
Abdul Samad, N.A.
Valthlanathan, T.
Aziz, S.M.
Brander, C.E.
Editors
Advisors
Journal Title
Journal ISSN
Volume Title
Type:
Conference paper
Citation
Proceedings APCCAS 2006, 2006, pp.674-677
Statement of Responsibility
Conference Name
2006 IEEE Asia Pacific Conference on Circuits and Systems (APCCAS 2006) (4 Dec 2006 - 7 Dec 2006 : Singapore)
Abstract
This paper presents the design and prototype development of a wireless receiver to supply power to biomedical devices used inside the human body. It is designed using the principle of magnetic induction to supply approximately 100 mW of power. Single coil and 3-coil orthogonal configurations are investigated, including simulations in SPICE. Prototype for a single coil receiver was developed and tested, where good correlation between simulation and experimental results was achieved. These initial experiments and prototype will form the basis for developing a battery-free, miniature power supply system for biomedical devices. ©2006 IEEE.
School/Discipline
Dissertation Note
Provenance
Description
Access Status
Rights
Copyright IEEE 2006