Removal of interferences in the speciation of chromium using an octopole reaction system in ion chromatography with inductively coupled plasma mass spectrometry

dc.contributor.authorChen, Z.
dc.contributor.authorMallavarapu, M.
dc.contributor.authorNaidu, R.
dc.date.issued2007
dc.description.abstractWhen inductively coupled plasma mass spectrometry (ICP-MS) is used for the detection of <sup>52</sup>Cr (83.8%) species in ion chromatography (IC), interference from polyatomic ions such as <sup>40</sup>Ar<sup>12</sup>C<sup>+</sup> and <sup>35</sup>Cl<sup>16</sup>O<sup>1</sup>H<sup>+</sup>, often occurs due to their mass at m/z 52. To address the issue, an octopole reaction system (ORS) in ICP-MS, including He and H<inf>2</inf> modes, was used to determine whether the background and interference from polyatomic ions could be reduced in the detection of <sup>52</sup>Cr. Compared to standard mode, the polyatomic ions were reduced by either He or H<inf>2</inf>; for example, more than 97 and 98% of <sup>35</sup>Cl was removed using a He and H<inf>2</inf> tune, respectively. However, the detection sensitivity for <sup>52</sup>Cr was decreased, for example, the sensitivity was 27.95 and 67.02% of the standard mode for Cr(EDTA)<inf>2</inf> <sup>-</sup>using a He and H<inf>2</inf> tune, respectively. The H<inf>2</inf> tune is recommended for the detection of <sup>52</sup>Cr at a flow rate of 2.0 mL/min with detection limits in the range of 0.2-0.4 μg/L. The developed method was used to measure chromium speciation in waters containing high concentration of chloride. © 2007 Elsevier B.V. All rights reserved.
dc.identifier.citationTalanta, 2007; 73(5):948-952
dc.identifier.doi10.1016/j.talanta.2007.05.030
dc.identifier.issn0039-9140
dc.identifier.urihttps://hdl.handle.net/1959.8/51692
dc.language.isoen
dc.publisherElsevier BV
dc.rightsCopyright status unknown
dc.source.urihttps://doi.org/10.1016/j.talanta.2007.05.030
dc.subjectchromium speciation
dc.subjectcollision reaction cell
dc.subjectcontaminated water
dc.subjectIC-ICPMS
dc.titleRemoval of interferences in the speciation of chromium using an octopole reaction system in ion chromatography with inductively coupled plasma mass spectrometry
dc.typeJournal article
pubs.publication-statusPublished
ror.mmsid9915911477001831

Files

Collections